  • AS950 All Weather Sampler Bundle with 1 – 5.5 Gallon Bottle

AS950 All Weather Sampler Bundle with 1 – 5.5 Gallon Bottle

Sampling has never been this easy.

The AS950 All-Weather Refrigerated Sampler simplifies sampling with a large, full-color display for intuitive, single-screen programming, USB upload and download capabilities and the ability to copy programs sampler to sampler to save time on tedious programing, and an error alert and status screen to eliminate uncertainty and simplify troubleshooting. The custom-designed air-sensing thermostat controls temperature in accordance with USEPA and international guidelines, preserving samples regardless of outside temperatures and conditions.

FEATURES
Easiest and Most Intuitive Operation

The large full color display and intuitive programming give you access to all your programmable criteria on a single screen – eliminating scrolling through menus and supporting error-free operation.

Most Convenient Data Transfer and Programming Available

The AS950 is the only sampler that utilizes a USB drive to upload and download data and copy programs from one sampler to another.

Confidence in Your Sampling Process

The program status screen instantly communicates alarms, missed samples and program progress for quick and easy troubleshooting.

Resists Corrosion

The All-Weather Refrigerated (AWR) sampler base is designed to endure humid and highly corrosive environments, minimizing damage caused by corrosive gases, rodents, and standing water to guarantee environmental integrity.

Accurate and Consistent Sample Preservations

The custom-designed air-sensing thermostat controls temperature in accordance with USEPA and international guidelines, preserving samples regardless of outside temperatures and conditions.

Specs
Alarm Configurable alarms that show on status screen and are recorded in diagnostics alarm logs. Alarms can be set for system diagnostics and logging such as program end, sample complete, missed samples and full bottle. Channel alarms are setpoint alarms for the recorded measurements (channels), such as pH, level and power supply voltage.
Automatic Shutdown Multiple Bottle Mode: After complete revolution of distributor arm (unless Continuous Mode is selected)
Composite Mode After preset number of samples have been delivered to composite container, from 1 to 999 samples, or upon full container.
Base Configuration 115V base with lock
Base Type No controller compartment heater
Body material IP24, low-density Polyethylene with UV inhibitor
Bottle Set Composite 5.5 gallon polyethylene bottle kit
Certifications CE, UL
Communication Capabilities USB and optional RS485 (Modbus)
Cooling system Top mounted compressor and fan-forced air cooled condenser; 11/5 HP 115 VAC: 115 °C (239 °F) thermal overload protector, 7.1 locked rotor amps; 230 VAC: 120 °C (248 °F) thermal overload protector, 7.6 A peak start current; 3-sided wraparound plate type evaporator; Rigid foam insulation: 3 in. (7.6cm) sides, 5 in. (12.7cm) top, 6 in. (15.2cm) bottom; Lockable lid to prevent tampering with programming; Recovery Time: Sampler temperature recovers to 4°C within 5 minutes after door has been held open for one minute in 24°C (75°F) ambient environment while in an active cooling cycle; Pull Down Time: from 24°C (75°F) to 4°C (39°F), 20 minutes; Temperature Control: 4 (±0.8) °C (39 (±1.5) °F)
Data logging SAMPLE HISTORY – Stores up to 4000 entries for sample time stamp, bottle number and sample status (success, bottle full, rinse error, user abort, distributor error, pump fault, purge fail, sample timeout, power fail and low main battery).
MEASUREMENTS – Stores up to 325,000 entries for selected measurement channels in accordance with the selected logging interval.
EVENTS – Ability to store up to 2000 entries in Sample History logging. Records Power On, Power Fail, Firmware Updated, Pump Fault, Distributor Arm Error, Low Memory Battery, Low Main Battery, User On, User Off, Program Started, Program Resumed, Program Halted, Program Completed, Grab Sample, Tube Change Required, Sensor Communication Errors, Cooling Failed, Heating Failed, Thermal Error Corrected.
Diagnostics View event and alarm logs as well as maintenance diagnostics
Dimensions
Please contact your Hach sales representative for details.
76 x 81 x 130 cm (30 x 32 x 51 in.)
NON-CONTACT LIQUID DETECT: Sample volume accuracy for applications that require complete tubing replacement
RAIN/RS485 PORT: Accepts Hach Rain Gauge (not included) or can be used as RS485 communications
TWO SENSOR PORTS: Accepts Hach digital Differential pH, Hach digital AV9000 analyzer with submerged area velocity flow and/or Hach digital US9000 ultrasonic level sensors
Display 1/4 VGA, Color; self-prompting/menu-driven program
Display Lock Function Access code protection prevents tampering
Housing Material PC/ABS blend, NEMA 4X, 6, IP68, corrosion and ice resistant (Controller)
Inputs One 0/4-20 mA input for flow pacing
Intake Strainers: Choice of PTFE and 316 stainless steel construction, or all 316 stainless steel in standard size, high velocity, and low profile for shallow depth applications
Purge: Air purged automatically before and after each sample;
duration automatically compensates for varying intake line lengths
Rinse: Intake line automatically rinsed with source liquid prior to each sample, from 1 to 3 rinses
Retries or Fault: Sample collection cycle automatically repeated from 1 to 3 times if sample not obtained on initial attempt
Internal Clock ±1 second per day at 25 °C (77 °F)
Manual Sampling Initiates a sample collection independent of program in progress
Memory Sample history: 4000 records;
Data log: 325,000 records;
Event log: 2000 records
Operating Mode Continuous or non-continuous
Pollution Degree 2
Power requirements (Hz) 50 – 60 Hz
Power requirements (Voltage) 115 V
Power supply 115V
Probes included No
Programs Dual Programs: Up to 2 sample programs can be run sequentially, in parallel, or according to day of week scheduling; enabling a single sampler to function like multiple samplers
Pump flow rate 4.8 L/min (1.25 gpm) at 1 m (3 ft) vertical lift with 3/8-in intake tube typical
Rain gauge input No
Refrigeration Components and Plumbing Corrosion protected with conformal all exposed copper tubing is insulated to avoid sweating and condensation
Sample Container 1 – 5.5 Gallon
Sample Interval Select. in single increments from 1 to 9,999 flow pulses or 1 to 9,999 mins in one minute increments
Sample Pump High-speed peristaltic, dual roller, with 0.95 ID x 0.16 OD cm (3/8 ID x 5/8-in. OD) pump tube; Pump body IP37, Polycarbonate cover
Sample Transport Velocity 0.9 m/s with 4.6 m vertical lift, 4.9 m of 3/8 inch vinyl intake tubing, 21 °C and 1524 m elevation
Sample Trigger When equipped with flow sensor or pH/temperature sensor or peripheral monitoring options, sampling can be triggered upon an upset condition when field selectable limits are exceeded
Sample Volume Accuracy Typical: ±5% of 200 mL sample volume with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of 3/8- in. vinyl intake tube, single bottle, full bottle shut-off at room temperature and 1524 m (5000 ft) elevation
Sample Volume Repeatability Typical: ±5% of 200 mL sample volume with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of 3/8- in vinyl intake tube, single bottle, full bottle shut-off at room temperature and 1524 m (5000 ft) elevation
Sampling Mode Pacing: Time Weighted, Flow Weighted, Time Table, Flow Table, Event
Distrib.: Single bottle composite, multi-bottle composite, multi-bottle discrete, bottles per sample, samples per bottle or a combination of bottles per sample and samples per bottle
Sensor Ports 1
Sensors None
Special Feature with Lock
Status Messages Communicates what program is running, if there are any missed samples, when the next sample will be taken, how many samples remain, number of logged channels, time of last measurement, memory available, number of active channels, if alarms were triggered, when alarms were triggered, active sensors and cabinet temperature
Suction Hose Outer Diameter 1/4″ and 3/8″ ID PTFE-lined Polyethylene with protective outer cover available in clear and black.
Tubing Pump tubing: 9.5 mm ID x 15.9 mm OD Silicone
Intake tubing: 1.0 – 4.75 m minimum length, 1/4″ or 3/8″ ID Vinyl or 3/8″ ID PTFE-lined Polyethylene with protective outer cover (black or clear)
User Interface
Membrane switch keypad with 2 multiple function soft keys
User Interface Languages:Chinese, English, French, German, Italian, Spanish, Portuguese
Vertical Lift 8.5 m using 8.8 m maximum of 3/8-in. vinyl intake tube at sea level at 20 to 25 °C
Warranty 12 months
Weight 86 kg
What’s included? 1 ea. AS950 controller on 115V All Weather base
1 ea. Bottle kit (AW010060) includes 1- 5.5 gallon poly bottle w/cap (6494), composite tube support (8838) and full bottle shut-off (8847)
1 ea. 25′ vinyl intake tubing (920)
1 ea. strainer (926)
Downloads

Brochure

Download

Manual

Download