Flurometers

Turner Designs

10 AU

Measurements are extremely stable with stability measured in months or weeks as opposed to days or hours. Watertight filter paddles enable quick change of excitation and emission filters so a variety of compounds can be easily measured on-site.

Optical kits with appropriate filters and lamps are available for a wide range of applications and can be easily installed in the field or lab. Individual filters and lamps can also be purchased for customized applications.

FEATURES
  • Exceptional Sensitivity
  • Automatic or Manual Ranging
  • Analog and Digital Output
  • Field Installable Optical Kits
  • Easy to Read LCD Display
Specs

10AU Detection Limits
Extracted Chlorophyll a 0.025 µg/L using EPA Method 445.0
Rhodamine WT Dye 0.01 ppb RWT in potable water
Fluorescein Dye 0.01 ppb Fluorescein in potable water


10AU Linear Range
Extracted Chlorophyll a  0 – 250 µg/L
Rhodamine WT Dye 0 – 250 ppb
Fluorescein Dye  0 – 250 ppb


10AU Electrical Specifications
Light Source  Lamp (4 watts, 8000 hours lamp life)
Detector  PhotoMultiplier Tube; Standard (300-650 nm), Red (185-870 nm)
AC Power  100-130 V; 200-240 V, 50/60 Hz, 30 watts
DC Power  11-16 V; 2.5 amperes
Digital Output  100% ASCII data output via RS-232 at 4800 or 9600 bps
Analog Output  0.1, 1, 2, or 5 volts


10AU Physical Specifications
Weight  12.6 kg (28.5 lbs)
Height 24 cm (9.45 in)
Width 55 cm (21.65 in)
Depth  34 cm (13.39 in)
Operating Temperature  -55° C, 32-131° F (ambient)