- Home
- Turner Designs
- Flurometers
- AquaFluor

AquaFluor
Its small size and durability make the AquaFluor® perfect for the user who needs quick measurements away from the laboratory. Dual-channel capability allows users to measure fluorescence and turbidity in one sample.
While small in size, performance has not been compromised. The AquaFluor® brings a new level of simplicity and economy to Turner Designs’ line of fluorescence-based instrumentation.
- FEATURES
-
Features:
- Dual channels allow for quick toggling between two applications
- Small size easily fits in shirt or jacket pocket
- Dustproof and watertight for field use, it even floats!
- Single-point plus blank calibration
- 12 bit resolution
- Operates using AAA batteries for >1,000 measurements per set
- 5 second warm up
- Low maintenance field fluorometer
- Specs
-
• Specifications
Application Minimum Detection Limit Linear Range Ammonium 0.1 μM 0 – 100 μM* CDOM/FDOM 0.1 ppb 0 – 1000 ppb Chl a Extracted – Acidification 0.5 μg/L 0 – 300 μg/L Chl in vivo 0.3 μg/L 0 – 300 μg/L Fluorescein Dye 0.4 ppb 0 – 400 ppb Optical Brighteners 0.5 ppm 0-30,000 ppm Phycocyanin
(Freshwater Cyanobacteria)150 cells/ml 0-150,000 cells/ml Phycoerythrin
(Marine Cyanobacteria)150 cells/ml 0-150,000 cells/ml Rhodamine Dye 0.4 ppb 0 – 400 ppb Turbidity 0.5 NTU 0 – 1000 NTU
*Ammonium linear range >10 μM requires minicell adapter and minicells
AquaFluor Physical Specifications Weight in Air 13.9 oz; 0.4 kg Size 1.75” x 3.5” x 7.25”; 4.45 cm x 8.9 cm x 18.4 cm Warm-up Time 5 seconds Case IP 67 standard; dustproof/waterproof Temperature 41-104°F; 5-40°C
AquaFluor Electrical Specifications Power 4 AAA batteries Detector Photodiode 300nm-1,000 nm capable - Downloads
-
Brochures brochure-20.pdf Download now Manuals Aquafluor-Manual.pdf Download now Additional files additional-15.pdf Download now