Handheld comms

Turner Designs

DataBank Logger

Data can be stored and downloaded to a PC. Sensor calibration is simplified through continuous sensor readings with results graphically displayed. Sensor power is provided directly from the DataBank which is powered by internal rechargeable batteries, an external 9-30VDC power source such as a 12V marine battery, or AC power. Available in two configurations: a handheld package for easy transport during mobile field studies and an enclosure for long-term stationary field deployment.

FEATURES
  • Stores up to 9,999 records and 16 calibrations
  • Includes auto-gain function
  • Simplifies sensor calibration
  • Station provides protection against corrosion, windblown dust and rain, water splash-down
  • Works with Cyclops-7™ Submersible Sensors
  • new Works with C-sense™ pCO2 Sensors
Specs

DataBank Electrical Specifications
Selectable Voltage: Software selectable 3 to 12 volts
Sensor Current: 150 mA at 3V to 100 mA at 12V
Parameters Warm-up Time: 0-25 seconds
Gain: 1-4
Inputs Input Selection: Switch; Differential Voltage, Differential Current or Dual Current
Range: 0-5V or 0-25 mA
Resolution: 12 bits
Sampling Rate: 1000 Hz (bursts); groups of bursts re-averaged @ 30 Hz
Outputs Two, Open Drain, up to 200 mA @ 15V max
(Used primarily for gain selection)
Storage Non-Volatile, 1280 Kb
(9,999 records, 1 record = sensor reading, date, time and location)
Current Draw Computer: < = 45 mA (at full charge). Sleeping < 20 uA
LCD Backlight: Full power ~ 30 mA. Low Power ~ 15 mA
Battery Type: NiMH rechargable, 4.8 volts, 4,000 mA/Hr.
Life: Continuous up to 70 hours (dependant on sensor/settings)
Note: When using the DataBank with the C-sense, it is recommended that the voltage be set to 6 and the warm-up time be set to 3 minutes.  Using these settings with a 5 minute sampling interval, you can expect a fully charged DataBank to collect 300 samples.
Charge Time ~3 hours (for a fully discharged battery; ambient @ 25 degrees C)
Communications USB


Handheld DataBank Physical Specifications
Size Main body: 4.6″W x 3.2″H, x 3.4″D, Handle: 4.8″ long
(117mm W x 81mm H x 86 mm D, Handle: 122mm)
Overall: 11”H (279 mm H) with handle, strain-relief and connector.
Weight 1.7 lb. (.76 kg) w/o pigtail


DataBank Station Physical Specifications
Size 8.27”W x 8.27”H x 5.12”D (210 mm W x 210 mm H x 130 mm D)
Weight 3.45 lb. (1.56 kg) w/o pigtail
Enclosure Approximates IP66 & NEMA 4x
Dessicant provides quick indication of moisture intrusion


Cyclops-7 Performance
Linearity 0.99R2


Application Minimum Detection Limit Dynamic Range
CDOM/FDOM 0.15 ppb**
0.5 ppb***
0-1250 ppb**
0-5000 ppb***
Chlorophyll in vivo
  Blue Excitation
Red Excitation
0.025 μg/L
0.5 μg/L
0-500 μg/L
>500 μg/L
Fluorescein Dye 0.01 ppb 0-500 ppb
Oil – Crude 0.2 ppb*** 0-2700 ppb***
Oil – Fine 10 ppb*
10 ppm****
>10,000 ppb*
>100 ppm****
Optical Brighteners 0.6 ppb*** 0-15,000 ppb***
Phycocyanin
(Freshwater Cyanobacteria)
2 ppbPC 0-40,000 ppbPC
Phycoerythrin
(Marine Cyanobacteria)
0.15 ppbPE 0-750 ppbPE
PTSA Dye 0.1 ppb*** 0-650 ppb***
Rhodamine Dye 0.01 ppb 0-1000 ppb
Tryptophan 3 ppb >20,000 ppb
Turbidity 0.05 NTU 0-3000 NTU


* 1,5 Napthalene Disulfonic Disodium Salt
** Quinine Sulfate
*** PTSA (1,3, 6, 8 – Pyrenetetrasulfonic Acid Tetrasodium Salt)
**** BTEX (Benzene, Toluene, Ethylbenzene, Xylenes)
PC Phycocyanin pigment from Prozyme diluted in Deionized water
PE Phycoerythrin pigment from Prozyme diluted in Deionized water

Cyclops-7 Physical Dimensions
Length x Diameter 5.7″ x 0.9″; 14.48 x 2.23 cm (SSt or Ti)
5.7″ x 1.25″, 14.48 x 3.18 cm (Delrin)
Weight 5.0 oz; 142 grams


Cyclops-7 Environmental Characteristics
Temperature Range Ambient: 0 to 50 deg C
Water Temp: -2 to 50 deg C
Depth Range 600 meters
Signal Output 0 – 5 VDC
Supply Voltage Range 3 – 15 VDC


Downloads

Brochure

Download

Manual

Download

Additional

Have Question About This Product?