Water Quality Sensors

Hydrolab

Nitrate Sensor

The Hydrolab Nitrate ISE Sensor (NO3-) is Factory integrated into the Hydrolab range of water quality meters:

• Hydrolab HL4
• Hydrolab HL7

 

HYDROLAB Nitrate Sensor: TECHNICAL SPECIFICATIONS

 

Parameters Measured
• Nitrate (NO3-)

Range
2 – 10,000 mg/L-N

Accuracy
+/- 0.1 5% or 2mg/L-N (larger value)

Resolution
0.01 mg/L-N

Operating Temperature
0°C to 40°C

Depth Range
• 15m

 

Hydrolab HL7

 

Note:  The Hydrolab Nitrate Sensor is an accessory and cannot be purchased as a standalone product.

The Hydrolab Nitrate Sensor is manufactured by Hydrolab and proudly represented by Aqualab Scientific.

FEATURES

Features:

  • The HYDROLAB Nitrate ISE sensor is a reference electrode immersed in a solution of fixed nitrate ion concentration separated by a membrane containing a chemical compound that reacts with the ion of interest, measuring electrical potential that varies with concentration.
Specs

Parameters Measured
• Nitrate (NO3-)

Range
2 – 10,000 mg/L-N

Accuracy
+/- 0.1 5% or 2mg/L-N (larger value)

Resolution
0.01 mg/L-N

Operating Temperature
0°C to 40°C

Depth Range
• 15m

 

Models

Available models include:

  • Hydrolab Nitrate Sensor (NO-3) for HL4 or HL7
    Product number# HLNO3
Downloads
Brochures
Hydrolab HL Series Brochure (AUS)

Hydrolab HL Series Brochure (AUS)

Hydrolab-HL-Series-Brochure-AUS.pdf
Download now
Manuals
Hydrolab HL Series Sensor Manual

Hydrolab HL Series Sensor Manual

Download now
Hydrolab HL Series ISE FAQ

Hydrolab HL Series ISE FAQ

Download now
Optional Accessories
  • Hydrolab Nitrate Calibration Solution (500ml) - Low Range
    Product number# 013800HY
  • Hydrolab Nitrate Calibration Solution (500ml) - High Range
    Product number# 013810HY
Videos
  • Hydrolab Nitrate Sensor: Maintenance & Calibration

Have A Question About This Product?